| Print |

DH-2000 Deuterium Tungsten Halogen Light Source

Part Number: 17002
Product Image: DH-2000 Deuterium Tungsten
Information:

Options and Accessories
Integrated shutters are also available with the DH-2000 and can be driven either by a switch or by a TTL signal. Another option is to include a filter holder with the system, which accepts filters up to 4 mm in thickness and as large as 25-mm square or 20-mm round in diameter. All versions of the DH-2000 have an SMA 905 Connector for easy coupling to our spectrometers and accessories via optical fiber, and a safety shutter for blocking the light when the fiber is not attached.
Adjustable Power
All versions of the DH-2000 have a potentiometer on the back of the light source to balance the light level between the deuterium and tungsten halogen light sources. This potentiometer allows you to adjust the optical power of the tungsten halogen light from 10-100%.
Optical Fibers
We recommend using our solarization-resistant optical fibers with all versions of the DH-2000.

 

DH-2000 Deuterium Light Source Specifications

Part Number

17002

Dimensions:

150 mm x 135 mm x 319 mm

Weight: 

3.5 kg

Power consumption: 

25 W (deuterium); 20 W (tungsten halogen)

Wavelength range: 

190-2500 nm (deep-UV deuterium and tungsten halogen bulbs)
215-2500 nm (standard deuterium and tungsten halogen bulbs)

Humidity: 

5-95% without condensation at 40 °C

Lamp current: 

Operating 85 V/0.3A

Lamp lifetime: 

1,000 hours

Lamp voltage: 

Ignition 580 V @20°C

Current voltage drift: 

<0.01% per hour

Current voltage stability: 

<5 x 10-6 peak-to-peak (0.1-10.0 Hz)

Operating temperature: 

5 °C - 35 °C

Power requirements: 

85-264 V 50/60 Hz

Radiation characteristic: 

Aperture 0.5 mm, numerical aperture 26° (13°); focused
Total power: 100 W

Power consumption: 

Approximately 78VA

Warm-up time: 

40 minutes (deuterium);
20 minutes (tungsten halogen)

Markings: 

CE; VDI/VDE 0160; EN 61010

 

DH-2000 Spectral Output

 

HL-2000 Spectral Output Graph

This spectral output graph shows the output of the DH-2000 (red) at full power and the DH-2000 with the tungsten halogen at low power (pink), D-2000 (blue). The height of the deuterium atomic emission line depends on the optical resolution of the spectrometer.